Cryogenic cathodoluminescence (cryo-CL)

Cryogenic cathodoluminescence (Cryo-CL) involves cooling samples to cryogenic temperatures, enhancing luminescence and enabling sharper spectral features by reducing thermal broadening. This technique stabilizes excitons, making it ideal for studying weakly-emitting samples or materials with fine spectral details.
MacBook mockup

Introduction

Cryo-CL is cathodoluminescence performed on a sample cooled to cryogenic temperatures, typically liquid nitrogen temperatures or below. Reducing the thermal agitation in the sample increases the luminescence intensity, enables the discrimination of narrow peaks within broad emission bands, and stabilizes exotic quasiparticles called excitons. Cryo-CL is thus most often used for samples which exhibit weak luminescence at room temperature, or which have fine spectroscopic features that are close in energy and therefore difficult to resolve.

Principles

Performing CL experiments at low temperature significantly impacts the obtained spectra. The main effects and mechanisms involved are discussed below.

Sharper spectral features

At low temperatures, spectral features in luminescence signals tend to become sharper as two distinct broadening mechanisms are reduced:

- Electronic thermal broadening

Thermal agitation at room temperature broadens the energy distribution of carriers (Figure 1) in a semiconductor. This in turn broadens the luminescence peaks involving a band edge as the carriers may recombine from a broad range of states. At cryogenic temperatures, the distribution of carriers become narrower as the Fermi-Dirac distribution approaches a step function.

A screenshot of a computer screenDescription automatically generated
Figure 1: Energy band diagram around the bandgap of an excited intrinsic semiconductor at steady-state showing the density of states g, the Fermi-Dirac distribution function f and the carrier density n, for electrons and holes. The quasi-Fermi levels EFc and EFv account for the non-equilibrium condition. 

- Phonon broadening

Electron-phonon interactions lead to peak broadening. Phonon broadening is reduced at cryogenic temperatures as the vibrational modes are less active.

It is noteworthy that cooling the sample does not reduce the width of all spectral features because of residual phonon broadening and other broadening mechanisms such as:

- Inhomogeneous broadening:

The presence of impurity, defect, alloy, or strain inhomogeneities in the material leads to variations in local emission energy and to a broadened overall peak.

 - Density-dependent broadening:

At higher densities, carriers can interact with each other, leading to broadening due to the spread in interaction energies.

Stronger luminescence signal

Lower temperatures reduce carrier-phonon interactions, which are responsible for non-radiative recombination processes. This increases the likelihood of radiative recombination, resulting in a stronger luminescence signal.

Higher exciton stability

After excitation, the electron and hole created in proximity attract each other and may combine in a hydrogen-like quasi-particle called an exciton. Free excitons move freely within the crystal. A free exciton may bind to a localized lattice defect or dopant, forming a bound exciton. When an electron and a hole that are paired within an exciton recombine, the recombination energy is reduced by the exciton binding energy. Consequently, excitons emit at energies a little below the band gap.

Free excitons whose binding energy is small compared to kT dissociate into free carriers. Likewise, weakly-bound excitons dissociate from their defect to become free excitons. As kT is reduced, excitons become more stable.

Instrumentation

The sample is typically cooled down to near liquid helium or liquid nitrogen temperature using a continuous-flow cryostat (Figure 2). The sample temperature is usually regulated using a resistive heater and a PID control loop.

A diagram of a wiring diagramDescription automatically generated with medium confidence
Figure 2: Schematic of a continuous-flow cryostat.

Example results

  1. The reduced thermal peak broadening at low temperature enables the discrimination of spectral features in GaN that are hardly visible at room temperature (Figure 3a).
  2. The enhanced luminescence signal at low temperature enables the detection of the weak emission of Si, an indirect bandgap semiconductor (Figure 2b). Peak attribution comes from Reference [1].
  3. The enhanced exciton stability enables the detection of the emission peaks from free excitons (FX) and bound excitons (BX) in diamond, as well as their phonon replica (Figure 2c) [2].
Figure 3: CL spectra of a) a GaN sample at room temperature (RT) and low temperature (LT), b) an Si sample at LT and c) a diamond sample at LT.

Benefits and Further Reading

Benefits and Further Reading

  • Enhanced luminescence for weakly-emitting samples
  • Sharper spectral features
  • Stabilization of both free and bound excitonic states

[1] P.J. Dean, J.R. Haynes, and W.F. Flood, Phys. Rev. 161, p. 711 (1967).

[2] N. Tappy, P. Gallo, A. Fontcuberta i Morral, C. Monachon, Carbon 191, p. 48 (2022).

References

Latest Scientific Publications

Synthesis, Characterisation, and Spectral Metrology of Plasmonic Titanium Nitride Films and Nanostructures. LIPINSKI, Arthur Filip PhD Thesis 2024
Applications

Cryogenic cathodoluminescence (cryo-CL)

Discover use cases involving Cryogenic cathodoluminescence (cryo-CL).

Text Link
Process development
Text Link
Materials Science
Text Link
Micro/Nanowire
Text Link
GaN device
Text Link
Life Science
Text Link
Power electronics
Text Link
Optoelectronics
Equipment

Our Cutting-Edge Equipment using Cryogenic cathodoluminescence (cryo-CL)

Explore our high-resolution cathodoluminescence tools designed for superior materials characterization.

Ready to revolutionize your materials characterization approach?

Get in touch with us today to discover how our state-of-the-art cathodoluminescence tools can elevate your research and industry applications.
Already more than 30 systems installed globally!